Desenvolvimento e teste de um dispositivo para estimulação torácica externa com função de marca-passo e avaliação da eficácia da estimulação cardíaca mecânica

The transcutaneous pacemaker has been used as therapy in cases of bradycardia with hemodynamic impairment, but it has important limitations, due to the high energy required to promote myocardial activation, causing discomfort and muscle stimulation. Thus, alternative forms of artificial cardiac s...

ver descrição completa

Na minha lista:
Detalhes bibliográficos
Autor principal: Oliveira Neto, Nestor Rodrigues de
Outros Autores: Fregonezi, Guilherme Augusto de Freitas
Formato: doctoralThesis
Idioma:pt_BR
Publicado em: Universidade Federal do Rio Grande do Norte
Assuntos:
Endereço do item:https://repositorio.ufrn.br/handle/123456789/57401
Tags: Adicionar Tag
Sem tags, seja o primeiro a adicionar uma tag!
id ri-123456789-57401
record_format dspace
institution Repositório Institucional
collection RI - UFRN
language pt_BR
topic Estimulação cardíaca artificial
Marca-passo
Acoplamento excitação-contração
Bloqueio atrioventricular
CNPQ::CIENCIAS BIOLOGICAS
spellingShingle Estimulação cardíaca artificial
Marca-passo
Acoplamento excitação-contração
Bloqueio atrioventricular
CNPQ::CIENCIAS BIOLOGICAS
Oliveira Neto, Nestor Rodrigues de
Desenvolvimento e teste de um dispositivo para estimulação torácica externa com função de marca-passo e avaliação da eficácia da estimulação cardíaca mecânica
description The transcutaneous pacemaker has been used as therapy in cases of bradycardia with hemodynamic impairment, but it has important limitations, due to the high energy required to promote myocardial activation, causing discomfort and muscle stimulation. Thus, alternative forms of artificial cardiac stimulation have been the object of studies. Mechanical stimulation is capable of causing cardiac activation and functioning as a pacemaker, similarly to activation by an electrical stimulus. A mechanical stimulus of a certain intensity can generate a mechanical deformation in the myocardium capable of initiating myocardial excitation, resulting in ventricular contraction. This work comprises two parts: the development and testing of a prototype electropneumatic device for transthoracic cardiac pacing in humans, and the performance of an experimental study aimed to assess the efficacy the of mechanical cardiac stimulation, comparing this form of stimulation with standard electrical stimulation, regarding cardiac activation, maintenance of ventricular command and induction of ventricular arrhythmia. Hearts from 10 adult toads (402 ± 101 g) of both sexes were partially exposed outside the body and randomly submitted to mechanical or electrical stimulation. Two pairs of copper electrodes were implanted to monitor and stimulate the heart. Mechanical stimulation was performed by means of a back-and-forth movement and impact of a 2 mm diameter metal rod on the external surface of the left ventricle (aiming at the lateral wall and apex). The electrical ventricular stimulation was provided by programmed temporary pacemaker generator in demand mode connected to two pacing wire leads. After performing the threshold measurements, each form of stimulation was maintained for five minutes, with an HR of 10 bpm above the baseline HR of each animal, maintaining the same energy and HR during each phase of the experiment. The electropneumatic device tested showed a linear correlation between the mechanical energy released and the pressure of the compressed air, providing increasing energy values within the predicted range for transthoracic cardiac pacing (considering a threshold of 0.04 to 1.5 J), reaching up to about 2.8J, with the maximum pressures provided by the hospital medical air network (average pressure of 5 bar). Data of the experimental study showed that direct mechanical stimulation (similarly to electric stimulation) can be maintained continuously by five minutes in toad hearts at low heart rate without loss of capture. Nevertheless, the maximum HR during mechanical stimulation before the loss of capture was close to baseline values and significantly lower than during electric stimulation in most toads, possibly because of different types of ion channels activated in mechanical and electrical stimulation. The occurrence of ventricular arrhythmias was similar in both forms of stimulation.
author2 Fregonezi, Guilherme Augusto de Freitas
author_facet Fregonezi, Guilherme Augusto de Freitas
Oliveira Neto, Nestor Rodrigues de
format doctoralThesis
author Oliveira Neto, Nestor Rodrigues de
author_sort Oliveira Neto, Nestor Rodrigues de
title Desenvolvimento e teste de um dispositivo para estimulação torácica externa com função de marca-passo e avaliação da eficácia da estimulação cardíaca mecânica
title_short Desenvolvimento e teste de um dispositivo para estimulação torácica externa com função de marca-passo e avaliação da eficácia da estimulação cardíaca mecânica
title_full Desenvolvimento e teste de um dispositivo para estimulação torácica externa com função de marca-passo e avaliação da eficácia da estimulação cardíaca mecânica
title_fullStr Desenvolvimento e teste de um dispositivo para estimulação torácica externa com função de marca-passo e avaliação da eficácia da estimulação cardíaca mecânica
title_full_unstemmed Desenvolvimento e teste de um dispositivo para estimulação torácica externa com função de marca-passo e avaliação da eficácia da estimulação cardíaca mecânica
title_sort desenvolvimento e teste de um dispositivo para estimulação torácica externa com função de marca-passo e avaliação da eficácia da estimulação cardíaca mecânica
publisher Universidade Federal do Rio Grande do Norte
publishDate 2024
url https://repositorio.ufrn.br/handle/123456789/57401
work_keys_str_mv AT oliveiranetonestorrodriguesde desenvolvimentoetestedeumdispositivoparaestimulacaotoracicaexternacomfuncaodemarcapassoeavaliacaodaeficaciadaestimulacaocardiacamecanica
_version_ 1790056219732344832
spelling ri-123456789-574012024-01-23T20:52:21Z Desenvolvimento e teste de um dispositivo para estimulação torácica externa com função de marca-passo e avaliação da eficácia da estimulação cardíaca mecânica Oliveira Neto, Nestor Rodrigues de Fregonezi, Guilherme Augusto de Freitas http://lattes.cnpq.br/3901786270968653 https://orcid.org/0000-0003-4938-7018 http://lattes.cnpq.br/2201375154363914 Fregonezi, Guilherme Augusto de Freitas Soares, Bruno Lobão https://orcid.org/0000-0003-4564-2288 http://lattes.cnpq.br/3124118595692286 Dias, Fernando Augusto Lavezzo Pinheiro, Francisco Irochima Belchior, Hindiael Aeraf Estimulação cardíaca artificial Marca-passo Acoplamento excitação-contração Bloqueio atrioventricular CNPQ::CIENCIAS BIOLOGICAS The transcutaneous pacemaker has been used as therapy in cases of bradycardia with hemodynamic impairment, but it has important limitations, due to the high energy required to promote myocardial activation, causing discomfort and muscle stimulation. Thus, alternative forms of artificial cardiac stimulation have been the object of studies. Mechanical stimulation is capable of causing cardiac activation and functioning as a pacemaker, similarly to activation by an electrical stimulus. A mechanical stimulus of a certain intensity can generate a mechanical deformation in the myocardium capable of initiating myocardial excitation, resulting in ventricular contraction. This work comprises two parts: the development and testing of a prototype electropneumatic device for transthoracic cardiac pacing in humans, and the performance of an experimental study aimed to assess the efficacy the of mechanical cardiac stimulation, comparing this form of stimulation with standard electrical stimulation, regarding cardiac activation, maintenance of ventricular command and induction of ventricular arrhythmia. Hearts from 10 adult toads (402 ± 101 g) of both sexes were partially exposed outside the body and randomly submitted to mechanical or electrical stimulation. Two pairs of copper electrodes were implanted to monitor and stimulate the heart. Mechanical stimulation was performed by means of a back-and-forth movement and impact of a 2 mm diameter metal rod on the external surface of the left ventricle (aiming at the lateral wall and apex). The electrical ventricular stimulation was provided by programmed temporary pacemaker generator in demand mode connected to two pacing wire leads. After performing the threshold measurements, each form of stimulation was maintained for five minutes, with an HR of 10 bpm above the baseline HR of each animal, maintaining the same energy and HR during each phase of the experiment. The electropneumatic device tested showed a linear correlation between the mechanical energy released and the pressure of the compressed air, providing increasing energy values within the predicted range for transthoracic cardiac pacing (considering a threshold of 0.04 to 1.5 J), reaching up to about 2.8J, with the maximum pressures provided by the hospital medical air network (average pressure of 5 bar). Data of the experimental study showed that direct mechanical stimulation (similarly to electric stimulation) can be maintained continuously by five minutes in toad hearts at low heart rate without loss of capture. Nevertheless, the maximum HR during mechanical stimulation before the loss of capture was close to baseline values and significantly lower than during electric stimulation in most toads, possibly because of different types of ion channels activated in mechanical and electrical stimulation. The occurrence of ventricular arrhythmias was similar in both forms of stimulation. O marca-passo transcutâneo tem sido usado como terapia em casos de bradicardia sintomática e sinais de má perfusão, porém apresenta limitações importantes, em virtude da alta energia necessária para promover a ativação miocárdica causa desconforto e estimulação muscular. Assim, formas alternativas de estimulação artificial tem sido objeto de estudos. A estimulação mecânica é capaz de provocar a ativação cardíaca e exercer função de marca-passo, de forma similar a ativação por um estímulo elétrico, uma vez que um estímulo de natureza mecânica, de certa intensidade, pode gerar uma deformação mecânica no miocárdio capaz de iniciar a excitação miocárdica, resultando na contração ventricular. Este trabalho compreende duas partes: o desenvolvimento e teste em bancada de protótipo de dispositivo eletropneumático para estimulação transtorácica cardíaca em humanos, com função de marca-passo cardíaco, com depósito de pedido de patente de invenção, e a realização de um estudo experimental em sapos (Rhinella jimi) para avaliar a eficácia da estimulação cardíaca mecânica, comparando esta forma de estimulação com a estimulação elétrica padrão, quanto à ativação cardíaca, manutenção da captura ventricular e indução de arritmia ventricular. A estimulação mecânica foi realizada por meio do movimento de vai-e-vem e impacto da uma haste metálica de 2 mm de diâmetro na superfície externa do ventrículo esquerdo (visando a parede lateral e o ápice), com uma frequência de disparo de 10 bpm acima da FC basal de cada animal e comparando esta forma de estimulação com a estimulação elétrica, em um modelo de corações de sapo in vivo que foi mantido parcialmente fora do corpo. A estimulação elétrica ventricular foi fornecida por gerador de marca-passo temporário programado em modo de demanda conectado a dois eletrodos. Após a realização das medidas dos limiares, cada forma de estimulação foi mantida por cinco minutos, com FC de 10 bpm acima da FC basal de cada animal, mantendo a mesma energia e FC durante cada fase do experimento. O dispositivo pneumático testado apresentou uma correlação linear entre a energia mecânica liberada e pressão do ar comprimido, fornecendo valores de energia crescentes dentro da faixa prevista para a estimulação cardíaca transtorácica (considerando um limiar de 0,04 a 1,5 J), podendo atingir de zero a cerca de 2,8J, com as pressões máximas fornecidas pela rede de ar comprimido hospitalar. Os achados do estudo experimental mostraram que a estimulação mecânica direta pode manter o comando ventricular, sem perda de captura, por um período de cinco minutos ou mais, em coração de sapo, quando a estimulação ventricular é realizada em baixa frequência cardíaca. No entanto, a FC máxima em que a estimulação mecânica mantém a captura ventricular 1:1 foi próxima a FC de repouso do animal e significativamente menor do que durante a estimulação elétrica, possivelmente devido a diferentes tipos de canais iônicos ativados em cada forma de estimulação. A ocorrência de arritmias ventriculares foi semelhante nas duas formas de estimulação. 2024-01-23T20:51:32Z 2024-01-23T20:51:32Z 2023-07-28 doctoralThesis OLIVEIRA NETO, Nestor Rodrigues de. Desenvolvimento e teste de um dispositivo para estimulação torácica externa com função de marca-passo e avaliação da eficácia da estimulação cardíaca mecânica. Orientador: Dr. Guilherme Augusto de Freitas Fregonezi. 2023. 129f. Tese (Doutorado em Biotecnologia) - Centro de Tecnologia, Universidade Federal do Rio Grande do Norte, Natal, 2023. https://repositorio.ufrn.br/handle/123456789/57401 pt_BR Acesso Aberto application/pdf Universidade Federal do Rio Grande do Norte Brasil UFRN PROGRAMA DE PÓS-GRADUAÇÃO EM BIOTECNOLOGIA